Analisis IQC dan Sigma Metrik Pemeriksaan Bilirubin Total Level Normal di RS Tipe C Kota Yogyakarta

DOI: https://doi.org/10.33650/trilogi.v7i3.16588
Authors

(1) * Nia Wulandari   (Universitas 'Aisyiyah Yogyakarta)  
        Indonesia
(2)  Havida Widyastuti   (Universitas 'Aisyiyah Yogyakarta)  
        Indonesia
(3)  Arifiani Agustin Amalia   (Universitas 'Aisyiyah Yogyakarta)  
        Indonesia
(*) Corresponding Author

Abstract


Background: The implementation of Internal Quality Control is crucial in clinical laboratories to ensure the accuracy and precision of test results, particularly for total bilirubin, which serves as an indicator of hepatobiliary disorders. Objective: This study aimed to analyze Internal Quality Control and sigma metrics for normal-level total bilirubin testing at a Type C hospital in Yogyakarta City by evaluating accuracy, precision, Levey-Jennings charts based on Westgard rules, and sigma metric values. Methods: This was a descriptive study with a quantitative approach utilizing secondary data from A total of 181 daily QC results from January–June 2025 were analyzed. Total sampling was employed. Data were processed using Microsoft Excel to calculate the mean, standard deviation (SD), Coefficient of Variation (CV%), bias (d%), and sigma metric values based on the 2024 CLIA Total Allowable Error (TEa) of 20%. Results: The Coefficient of Variation (CV%) values from January to June were 16.66%, 9.80%, 11.76%, 5.88%, 13.72%, and 10.78%, respectively. Only April met the precision criterion (CV ≤7%). Bias (d%) values were 8.62%, 9.80%, 8.82%, 10.59%, 11.76%, and 7.84%, respectively; all values fell below the Total Allowable Error (TEa) limit, indicating acceptable testing accuracy. Levey-Jennings charts revealed violations of the Westgard 1₂s rule from February to June and the 1₃s rule in May. Sigma metric values from January to June were 0.72, 1.04, 0.95, 1.56, 0.60, and 0.85, respectively. All results fell into the <3 sigma category. Conclusion: Total bilirubin testing demonstrated acceptable accuracy, yet precision was suboptimal. Evaluation of Levey-Jennings charts and sigma metric values indicates that the Internal Quality Control process requires improvement.


Keywords

Westgard rules; Total Bilirubin; Levey-Jennings; Internal Quality Control; Sigma metrics.







References


Adiga. (2021). Analisis Quality Control (Qc) Pemeriksaan Albumin Dan Bilirubin Menggunakan Kontrol Harian Dan Matriks Sigma. Borneo Journal Of Medical Laboratory Technology, 7(1), 481–489. https://doi.org/10.33084/bjmlt.v7i1.7863

Aisy, D. R., Hadi, W. S., & Shafriani, N. R. (2024). Analisis kontrol kualitas hemoglobin hematologi analyzer puskesmas x menurut grafik levey-jennings dan six sigma. Prosiding Seminar Nasional Penelitian Dan Pengabdian Kepada Masyarakat LPPM Universitas’ Aisyiyah Yogyakarta, 2, 1457–1465.

Bishop, M. L., Fody, E. P., & Schoeff, L. E. (2021). Clinical Chemistry: Principles, Techniques, and Correlations (8th ed.). Philadelphia: Wolters Kluwer.

CLIA. (2024). Clinical Laboratory Improvement Amendments (CLIA) Proficiency Testing Limits: Total Allowable Error (TEa) Table Standards. Centers for Medicare & Medicaid Services.

Erdani. (2024). Gambaran Kadar Bilirubin Sebelum Dan Sesudah Fototerapi Pada Bayi Baru Lahir Di Rumah Sakit Urip Sumoharjo Tahun 2023. Poltekkes Kemenkes Tanjungkarang.

Geto, Z., Getahun, T., Lejisa, T., Tolcha, Y., Bikila, D., Bashea, C., Meles, M., Habtu, W., Ashebir, G., & Negasa, B. (2022). Evaluation of sigma metrics and Westgard rule selection and implementation of internal quality control in clinical chemistry reference laboratory, Ethiopian Public Health Institute. Indian Journal of Clinical Biochemistry, 37(3), 285–293. https://doi.org/10.1007/s12291-021-00994-x

Hulzebos, C. V., Dijk, P. H., & van Imhoff, D. E. (2024). Management of neonatal hyperbilirubinemia: Current standards and future perspectives. Journal of Perinatology, 44(2), 123–131.

Kementerian Kesehatan Republik Indonesia. (2022). Peraturan Menteri Kesehatan Republik Indonesia Nomor 43 Tahun 2022 tentang Cara Penyelenggaraan Laboratorium Klinik yang Baik. Jakarta: Kementerian Kesehatan Republik Indonesia.

Kementerian Kesehatan Republik Indonesia. (2023). Peraturan Menteri Kesehatan Republik Indonesia Nomor 3 Tahun 2020 tentang Klasifikasi dan Perizinan Rumah Sakit.

Kesuma, R. A., Lestari, D., & Pratama, M. A. (2023). Pengaruh penanganan spesimen dan internal quality control terhadap hasil pemeriksaan bilirubin total. Jurnal Teknologi Laboratorium Medik, 12(1), 45–52.

Organisasi Internasional untuk Standardisasi. (2022). ISO 15189:2022 Medical Press.

laboratories — Requirements for quality and competence. Geneva: ISO.

Praja, R. W. (2022). Analisis Hasil Pengendalian Mutu Internal Pemeriksaan Kimia Klinik Dengan Metode Six Sigma Di Laboratorium Patologi Klinik Rsud Dr. H. Abdul Moeloek Provinsi Lampung Tahun 2021. Poltekkes Tanjungkarang.

Praptomo, A. J., Anam, K., & Raudah, S. (2021). Pengendalian Mutu Laboratorium Medis. Yogyakarta: Deepublish.

Rahmawati, A. M., Rinaldi, F. S., Riyani, A., & Kurnaeni, N. (2025). Evaluation and follow‑up of internal quality control of total bilirubin testing at RSIA “X”. INDOGENIUS, 4(3), 689–693. https://doi.org/10.56359/igj.v4i3.609

Rahmani, A. R., Amalia, A. A., & Novalina, D. (2024). Analisis Quality Control (QC) Pemeriksaan Albumin Dan Bilirubin Menggunakan Kontrol Harian Dan Matriks Sigma. Borneo Journal of Medical Laboratory Technology, 7(1), 481–489. https://doi.org/10.33084/bjmlt.v7i1.7863

Sari, D. P., Wibowo, A., & Handayani, S. (2023). Pengaruh paparan cahaya dan kondisi spesimen terhadap hasil pemeriksaan bilirubin total. Jurnal Patologi Klinik Indonesia, 10(3), 157–164.

Vyakarana S, Nori SNS, Rehan AR, Pothula Y, Yasam L, Agiwal V. Six sigma matrix and Quality Goal Index ratio in improving the quality of analytical phase in a clinical laboratory. Journal of Laboratory Physicians. 2024;16(4):515–521. https://doi.org/10.25259/JLP_56_2024

Westgard, J. O., & Westgard, S. A. (2021). Basic Method Validation: Training in Analytical Quality Management for Healthcare

Laboratories (4th ed.). Madison: Westgard QC Press.

World Health Organization. (2022). Global hepatitis report 2022. Geneva: WHO


Dimensions, PlumX, and Google Scholar Metrics

10.33650/trilogi.v7i3.16588


Refbacks

  • There are currently no refbacks.


Copyright (c) 2026 Nia Wulandari

Creative Commons License
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.

Creative Commons License
 

TRILOGI: Jurnal Ilmu Teknologi, Kesehatan, dan Humaniora
Published by Lembaga Penerbitan, Penelitian dan Pengabdian kepada Masyarakat (LP3M) of Nurul Jadid University, Probolinggo, East Java, Indonesia.